VetScan® Classic Installation Guide
Your Operator's Manual is the most complete, essential reference for the VetScan Chemistry Analyzer. Always refer to the Operator's Manual for detailed instructions and troubleshooting guidance. Place the manual in a convenient place for easy access.
Newsletters, product and manual updates will be automatically mailed to you and should be kept in your Operator's Manual.
Following is the simple, 3-step installation process.
Step 1: Unpack the analyzer and accessories.
- Place the box of reagent rotors in the refrigerator (2°-8° C, 36°-46° F) immediately upon arrival
- Remove analyzer from box and place on level surface
- Complete and mail warranty card
Step 2: Set up analyzer and electrical connections.
- Attach AC supply cord (with electrical wall plug) to power adaptor
- Attach DC power supply cord to back of VetScan
- Plug AC supply cord into grounded electrical outlet
- Make sure all connections are secure
NOTE: To prevent power surges or drains, do not plug the VetScan into the same circuit as your centrifuge, refrigerator, or any other high current device. Abaxis® recommends that the VetScan be plugged into a surge protector designed for use with a computer. Abaxis recommends the use of an uninterruptible power supply (UPS) if the VetScan will be used in an area prone to electrical surges or power outages.
The VetScan analyzer will automatically turn on when plugged in to the electrical outlet. The analyzer display will show the following messages:
"PERFORMING SELF-TEST"
"HEATING ROTOR CHAMBER TO OPERATING TEMPERATURE"
"OPEN DRAWER TO RUN A ROTOR"
You will need to push the OPEN/CLOSE key on the keypad to open the drawer.
Troubleshooting
Are you experiencing any of these problems?
- Analyzer does not turn on. (Verify that all connectors are firmly inserted and the electric outlet is working properly.)
- Drawer fails to open when you push the OPEN/CLOSE key on the keypad.
- Error messages pertaining to hardware or software are shown on the display.
Refer to Section 7, Troubleshooting in your Operator's Manual. If the Troubleshooting guide does not correct your problem, call Technical Support at 800-822-2947.
Step 3: Setting System Parameters
The following procedures are performed using the MENU key. The drawer must be closed in order to access these special functions.
1. Set time and date. See Operator's Manual, Section 4.3, "Changing Date and Time" for details. The analyzer is pre-set to Pacific time. NOTE: Date and time must be accurately set. The analyzer uses this information to determine reagent dating when reading the bar-coded rotor prior to sample analysis. Incorrect settings could cause an error message to be displayed.
EUROPEAN CUSTOMERS: If desired, set time and/or date formats. See pages 4-9 and 4-10 of the Operator's Manual.
2. Measurement units. The analyzer is pre-set to report results in common units (e.g., mg/dL). You may choose to change the units to SI units (e.g., mmol/L). To change units see Operator's Manual, Section 4.6, "Selecting Units."
If at any time, the normal sequence of events does not occur, refer to your Operator's Manual or call our Technical Support Department (800-822-2947). You can also email our Technical Support Department at technicalsupport@abaxis.com.
The Technical Support representative will ask for
- a description of the problem
- the analyzer serial number and
- software version number
Serial number and software version numbers can be accessed through the MENU key. A complete description can be found in the Operator's Manual, Section 4.7, "Viewing Analyzer Identification."
Congratulations! You have successfully installed the VetScan analyzer and are now ready to run samples.